BARNSTEAD|THERMOLYNENANOpure®DIamond™Analyticalultrapure water systemOperation ManualSeries 1190Model No. VoltageD11901 100V - 240VLT1190X1 • 6/22/00
Other AccessoriesOptional accessories for this unit include a float switch,low pressure switch, N.I.S.T. Traceable CalibrationModule, remote dispenser
Installing the Control Panel in aRemote LocationFor your convenience, the control panel can be removedfrom the unit and mounted at a convenient locati
12Water ConnectionsFeed Water Connection1. Locate the length of 3/8” (.95 cm) O.D. tubingprovided with the unit with a quick disconnectinsert on one e
13Main Power SwitchThe main power switch on the NANOpure DIamondAnalytical is located on the lower back right side of theunit (as you face the front o
SwitchesWhen the main power switch (on the lower back right sideof the unit as you face the unit) is ON, the switches onthe control panel function as
15Cartridge packs will come bagged with four manifold connection caps.Part ApplicationDIamond Kit Organic Free R/O & Distilled FeedD50280 Ultra-Lo
16Initial Rinse1. Attach the feed water line to the unit by snapping the quick disconnect coupling into the quick disconnect body in the lower left ba
17System Sanitization ProcedureYour NANOpure DIamond Analytical has been shippedwith a cleaning syringe (Cat. No. CMX25). When the sys-tem is new or w
1813. Press the UP or DOWN arrow to select (YES)and press ENTER. Display will now read, “UnitSanitizing.”14. Unit may now be left unattended until dis
191. Turn the system power ON by depressing themain power switch to the “I” position.2 The system greeting display indicates the typeof unit you own.
Safety Information ...
Adjusting Display Brightness1. From the (Idle) mode, when the display reads,“NANOpure Diamond (Idle) xx:xx:xxxx/xx/xx,” press the UP or DOWN arrow unt
212. Press ENTER.3. Press ENTER when the unit displays “PrinterAttached?”4. When prompted with “Date Format” press theUP or DOWN arrow to show the dat
222. Unlatch and open the right side door.3. Find and copy down the two cell constantsattached to the gray cable,leaving the electron-ics board mount
23Use of Standby ModeAt the end of the work day, place the NANOpureDIamond Analytical in Standby mode for the night. Toplace the unit in Standby mode:
247. Press ENTER.8. Press BACK to return to the (Idle) mode.The second line of the display will show a “Below Setpoint” message once every 8 to 10 sec
25USER SETTINGSSetting Volumetric Dispensing1. From the normal recirculating mode, when thedisplay is showing purity, press the UP orDOWN arrow until
26NotePress the STOP/STOP switch to can-cel the flush.Setting Timed Dispensing1. From the normal recirculating mode, when the dis-play is showing puri
272. Press the UP or DOWN arrow to select “MΩ-cm”or “uS/cm” Press ENTER.3. Press BACK to return to the normal recirculatingmode.Temperature Compensati
28Accessories D8964 (float switch) and D2706 (pressureswitch) are designed to protect the NANOpure DIamondAnalytical pump by alerting the system of an
29Performing an ElectronicCalibration Using the OptionalN.I.S.T. Calibration ModuleIf you purchased the optional N.I.S.T. calibration module(Catalog N
Your Barnstead|Thermolyne NANOpure DIamondAnalytical ultrapure water system has been designed withfunction, reliability, and safety in mind. It is you
30N.I.S.T. CALIBRATION MODULE11. Press BACK to return to the (Idle) mode.12. Turn the main power switch to the OFF “O” position. Unplug the unit.13. R
Connecting NANOpureDIamond to Computer andCommunicating Though the RS-232 Port UsingHyperterminal or ProcommHyperterminal RS-232 Capture Instructions1
32capture text. Accepting default will put the filewhich you name in the c:/windows directorycalled capture.txt and the file will be an ASCIItext file
3312. Type in 21 <Enter>, sets to COM2: or (Type in20 <Enter>, sets to COM1:), depending on yourcomputer.13. Type in 24 <Enter>, sav
347. From the “Printer Attached” display, press theUp or DOWN arrow for (YES).8. Press ENTER. Press BACK twice to return tothe (Idle) mode.9. During n
35down; or—the paper may be jammed in the printer. Turnthe printer OFF before checking for jammed paper.Remove the paper jam and turn the printer back
36General Cleaning InstructionsDisconnect electrical service to the unit. Wipe exteriorsurfaces with lightly dampened cloth containing mild soapsoluti
System SanitizationAfter sanitization is complete (approximately 3.5 hours),place the unit into its normal (recirculation) mode to allowthe system to
3814. Unit may now be left unattended until the display reads, “Sanitization Complete: PressENTER.”15. After sanitization is complete and the userpres
39Cartridge Pack ReplacementThe frequency with which you will need to clean your unitand replace your cartridge pack is dependent on yourfeed water’s
3. Avoid splashing disinfecting solutions on clothing or skin.4. Ensure all piping connections are tight to avoidchemical leakage.5. Ensure adequate v
0.2 Micron Filter ReplacementReplace the 0.2 micron final filter whenever any of the fol-lowing conditions occur: every 30 days, the product waterflow
41Fuse Replacement1. Turn off the NANOpure DIamond Analytical anddisconnect it from the power supply by removingthe power cord directly below the main
42Problem Possible Causes SolutionsNANOpure DIamond completely No electrical power to NANOpure Ensure NANOpure DIamond power inactive (pump and displ
43Poor quality feed water. If a Barnstead ROpure is the feed water source, check that the membrane is functioning properly.If a Barnstead Still is the
44Recommended SparesConsumablesConsumable parts are those required to support the day-to-day operation of this equipment.Barnstead|Thermolyne establis
45Safety StockFor critical applications where performance with minimum downtime is required, Barnstead|Thermolyne rec-ommends that you maintain a loca
46Flow ChartUnit Initialization & MeterCalibration CheckNANOpure DIamond (Idle)xx :xx:xx xx /xx/xxstart_stopkey pressedup or downkeys pressedSani
47Wiring DiagramCONNECTORREMOTE DISPLAYCSW1190X2SW1190X2SW1190X2(SPG)230V: (2) FSX54& FUSES:120V: (2) FSX47& D11931 ONLYMODEL #s D11911 LINENE
48Ordering ProceduresPlease refer to the Specification Plate located inside theleft side door for the complete model number, serial num-ber, and serie
49Barnstead|Thermolyne Corporation warrants that if a product manufactured by Barnstead|Thermolyne andsold by it within the continental United States
Congratulations on your purchase of a Barnstead|Thermolyne NANOpure DIamond Analytical ultrapurewater system. This water purification system is design
9001ISOREGISTEREDBarnstead|Thermolyne2555 Kerper Blvd.P.O. Box 797Dubuque, IA 52004-2241Phone: 319-556-2241 800-553-0039Fax: 319-589-0516E-Mail Addr
Dimensions and Clearance RequirementsDimensions 13.4” W x 19.5” H x 16.9” D (34.0 x 49.5 x 42.8 cm)ClearancesSides - 9” (22.9 cm) minimum for servicin
Environmental Conditions Operating: 4°C - 49°C; 20% to 80% relative humidity, non-condensing. Installation Category II (over-voltage) in accordance w
1. Remove the unit from its shipping container.Remove all contents carefully. Ensure that thefeed and drain tubing, sanitization cartridge,wall bracke
9Choosing a SiteThe NANOpure DIamond Analytical features a removablecontrol panel display which allows the system to bemounted almost anywhere within