1LT1056X2• 9/24/98BARNSTEAD|THERMOLYNE CORPORATIONEASYpure™ UV/UFOPERATION MANUALAND PARTS LISTSeries 1056

Seite 2 - Table of Contents

10you. Viewed this way, the product connector isthe one on the left. The reject connector is theone on the right. (See Figure 5) Remove the nut,stainl

Seite 3 - Safety Information

117. Locate the UV chamber. Carefully remove theblack plastic cover from the UV chamber bypulling the cover straight up. Do not pull onthe cable.8. Re

Seite 4 - SAFETY INFORMATION

12Wall Mounting (Optional)Install the optional wall bracket on the wall in a location thatis accessible to water, electricity and an atmosphericallyve

Seite 5 - Introduction

13Reject Water ConnectionWhen the EASYpure UV/UF flushes its membrane, thewater used is sent to drain through this connection. Toinstall:1. Locate the

Seite 6 - General Specifications

14Main Power SwitchThe main power switch on the EASYpure UV/UF is located onthe back right of the unit (as you face the front of the unit),directly ab

Seite 7 - Declaration of Conformity

15Pressing the Flush switch twice will initiate a five minuteflush of the ultrafilter. During this five minute flush, wateris sent to drain through th

Seite 8 - Unpacking and Installation

16Run ModeSince not all qualities of permissible feedwater will reachmaximum resistivity after one pass through the unit’scartridges (especially as th

Seite 9 - Installation

17Initial SanitizationYour EASYpure UV/UF has been shipped with a sanitiza-tion cartridge (Catalog Number D50245) and two emptycartridges to allow you

Seite 10 - INSTALLATION

188. Open the draw-off valve to depressurize thesystem. Leave valve open until solutioncompletely stops flowing, draining as muchof the system as poss

Seite 11 - Bench Mounting

1910. Press the “Flush” button twice to begin anextended flush. When water begins exiting thesystem, close the draw-off valve.11. When the extended fl

Seite 12 - Water Connections

2Table of ContentsSafety Information ...

Seite 13 - Reject Water Connection

201. Press the START/STOP button to start. TheEASYpure UV/UF’s pump will begin to run andthe Purity meter will display the resistivity of thewater in

Seite 14 - Controls

21Installing Float or PressureSwitchAccessories D0606 (float switch) and D2706 (pressureswitch) are designed to protect the EASYpure UV/UFpump by aler

Seite 15 - CONTROLS

22System Sanitization and CartridgeReplacementThe frequency with which you will need to clean your unit andreplace cartridges is dependent on your fee

Seite 16 - Operational Modes

238. Repeat steps 2 - 6 with the two empty car-tridges supplied with your EASYpure UV/UF,placing them in the center and left-handpositions.9. Close th

Seite 17 - Initial Operation

24Ultrafilter ReplacementThe length of your ultrafilter’s life will depend to someextent on conditions and use in your lab. When you areunable to comp

Seite 18 - Up (Refer to Figure 10)

25necessitating the EASYpure UV/UF be placed in a recircu-lating mode for up to one hour to achieve organic content of2 ppb TOC.To replace the ultravi

Seite 19 - INITIAL OPERATION

2612. Reattach the power cord and the pump protectoror the pump interlock and reconnect the unit tothe power supply.13. Operate normally.Fuse Replacem

Seite 20 - Normal Operation

27UV Solid State Ballast FuseReplacement1. Turn off the EASYpure UV/UF and disconnectit from the power supply. Remove the powercord and the pump prote

Seite 21 - Installing Float or Pressure

282. Depressurize the system by opening thedraw-off valve, allowing water to drain untilno more flows from the valve.3. Open the cartridge access door

Seite 22 - Maintenance and Servicing

2916. Reinstall the screw holding the cell-cable retain-ing clip. Reinstall membrane switch lead. Re-place the front cover. Replace the Luer fitting w

Seite 23 - 0.2 Micron Filter Replacement

3Safety InformationYour Barnstead EASYpure™ UV/UF has been designedwith function, reliability, and safety in mind. It is yourresponsibility to install

Seite 24 - UV Lamp Replacement

30TroubleshootingProblem Possible Causes SolutionsEASYpure UV/UF completely No electrical power to Ensure that the EASYpureinactive. (Pump not operati

Seite 25 - MAINTENANCE AND SERVICING

31TROUBLESHOOTINGProblem Possible Causes SolutionsRecirculated water will not rinse Exhausted cartridge Replace the cartridge asup to desired purity l

Seite 26 - Fuse Replacement

32Problem Possible Causes SolutionsIf a Barnstead Still is theShort cartridge life (cont.) feedwater source, ensure that themembrane is functioning pr

Seite 27 - Cleaning the Resistivity Cell

33ConsumablesConsumable parts are those required to support the day-to-day operation of this equipment.Barnstead|Thermolyne establishes two types of c

Seite 28

34General MaintenanceGeneral maintenance parts are defined as laboratory level repair parts which do not require great expertise orspecial tools for i

Seite 29 - Shutdown

35100 Volt Wiring SchematicWiring DiagramsNANONANOGNDL2L1PUMPPRODFEEDPUMPFEEDPRODBLACK100 VAC, 50/60 Hz3.0 ALINEGROUND1FILTERLINE1PUMP INTERLOCKWITH J

Seite 30 - Troubleshooting

36120 Volt Wiring SchematicWIRING DIAGRAMSGND120 VAC, 50/60 HzGROUNDLINE1LINEFILTER1PUMP INTERLOCKWITH JUMPER PLUGFOR USE WITHOPTIONAL PRESSUREOR FLOA

Seite 31 - TROUBLESHOOTING

37230 - 240 Volt Wiring SchematicWIRING DIAGRAMSGND240 VAC, 50/60 HzGROUNDLINELINEFILTER1PUMP INTERLOCKWITH JUMPER PLUGFOR USE WITHOPTIONAL PRESSUREOR

Seite 32

38Exploded Views

Seite 34 - Safety Stock

42. This device is to be used with water feeds only.Sanitizing/cleaning agents must be used incompliance with instructions in this manual.Failure to c

Seite 35 - Wiring Diagrams

40Ordering ProceduresPlease refer to the Specification Plate for the complete model number, serial number, and series number whenrequesting service, r

Seite 36 - WIRING DIAGRAMS

41Barnstead|Thermolyne Corporation warrants that if a product manufactured by Barnstead|Thermolyne andsold by it within the continental United States

Seite 39 - EXPLODED VIEWS

449001ISOREGISTEREDBarnstead|Thermolyne2555 Kerper Blvd.P.O. Box 797Dubuque, IA 52004-2241Phone: 319-556-2241 800-553-0039Fax: 319-589-0516E-Mail Ad

Seite 40 - Ordering Procedures

5The Barnstead EASYpure UV/UF is a line-fed waterpurification system designed to provide pyrogen-free, lowTOC reagent grade water that exceeds ASTM, T

Seite 41 - One Year Limited Warranty

6Dimensions and Clearance RequirementsEASYpure UV/UF dimensions: 12 " W X 19" D X 17 1/8" H (30.5 cm X 48.3 cm X 43.5 cm).Clearances:Si

Seite 42

7Environmental ConditionsOperating: 4°C - 49°C; 20% - 80% relative humidity,non-condensing. Installation Category II(over-voltage) in accordance with

Seite 43

8Unpacking1. Remove the unit from its shipping container.Ensure that the ultrafilter, UV lamp, feed and rejecttubing, final filter, sanitization cartr

Seite 44 - Barnstead

93. Place the grab ring and back-up ring in the hexnut in the order and orientation shown in Figure3. Thread the nut onto the adapter. DO NOT usethe o