LT718X1 • 12/4/00 Serial Number: _______________________________ThermolyneBarnstead|Thermolyne CorporationType 30400Thermolyne FurnaceOPERATION MANUAL
10Observe These Warnings Before Operating Your Furnace: WARNINGDo not use in the presence of flammable or combustible chemicals; fire or explosion may
11 Barnstead |Thermolyne • 800-553-0039 NOTEWhen performing the operations with the controller, remember that if more than eight to tenseconds elapse
12 NOTEIf the power to furnace is "turned off" or interrupted while in "tuning,"upon returning power tofurnace, the controller dis
13 Barnstead |Thermolyne • 800-553-0039Operation of 2 Ramp & 2 Dwell ProgrammableModels (F30420C-70, F30428C-70, F30420C-60-70,F30428C-60-70)Contr
14Cycle Indicator:The amber cycle light will illuminate whenever the door is closed and the actual chamber temperature isless than the setpoint temper
15 Barnstead |Thermolyne • 800-553-0039The next three parameters - Proportional Band, Integral and Derivatives - are the three control parametersof a
16Operating the Controller CAUTIONDo not exceed limitations for continuous or intermittent operating temperature shown inthe General Specifications (i
17 Barnstead |Thermolyne • 800-553-0039Push "PAR" until "SP" is displayed. Push "UP" or "DOWN" buttons and mod
18controller flashes and "Hold" is indicated at the "prog" parameter, the controller is in Hold. Now you areable to adjust all par
19 Barnstead |Thermolyne • 800-553-00398 Ramp & 8 Dwell Multi-Programmable Models(Models F30430CM, F30438CM, F30430CM-60,F30438CM-60)Controls and
2Read All Warnings, Cautions and Instructions Carefully Before Operating This Thermolyne Furnace.IMPORTANT INFORMATIONThis manual contains important o
20 WARNINGTo avoid electrical shock, this furnace must have the door switch connected and operatingproperly. If the furnace power light does not go ou
21 Barnstead |Thermolyne • 800-553-0039The next three parameters - Proportional (PB) Integral (+i) and Derivative (+d) - are the three controlparamete
22be running a multi-step program, set the furnace's setpoint to the value of PL1 (Program Level #1).)At the end of the minute, the "SP"
23 Barnstead |Thermolyne • 800-553-0039Program EntryTo Link Programs TogetherPush scroll button until "Cnt n" is displayed. Press and releas
24achieved. By pushing the up or down button the number of loops can be set at any value from 1 to999.Set the Holdback Feature5. Scroll through the ma
25 Barnstead |Thermolyne • 800-553-0039When the reset has been enabled, the parts of the display associated with programming will beextinguished and t
26Exhaust tubing hook-up. NOTEAppropriate exhaust must be provided to remove smoke and gases produced in an ashingprocedure.Using accessory tubing ava
27 Barnstead |Thermolyne • 800-553-0039Problem SolvingThe Problem Solving section is intended to aid in defining and correcting possible service probl
28Problem Probable Causes Corrective ActionControl/Programmer locked up,not allowing any change tovalues.Program is running. To cancel program, press
29 Barnstead |Thermolyne • 800-553-0039MSDS’s, or additional information concerning the handling of refractory ceramic products, please contactthe Cus
3 Barnstead |Thermolyne • 800-553-0039To Replace a Platinel II Thermocouple...
30To Replace Solid State Relay:a. Disconnect furnace from power supply.b. Remove the screws on the front dial and the screws on the lower back cover.c
31 Barnstead |Thermolyne • 800-553-0039d. With the door closed loosen the screws on the microswitch and slide the switch downward, so thatthe screws a
32Replacement Parts ListSeries 718Part Number Description Models Used OnCN71X55 Control unit, automatic model F30420C,F30428CCN71X40 Control Unit, Pro
33 Barnstead |Thermolyne • 800-553-0039Wiring Diagrams Series 718Models F30430CM, F30438CM, F30430CM-60 & F30438CM-60
34Wiring Diagrams Series 718Models F30420C-70, F30428C-70, F30420C-60-70, F30428C-60-70
35 Barnstead |Thermolyne • 800-553-0039Wiring Diagrams Series 718Models F30420C & F30428C
36ThermalCeramicsMaterial SafetyData SheetDate Revised: 7/2/91PRODUCT IDENTIFICATIONTrade Name(s): CERAFIBERGeneric Name: REFRACTORY CERAMIC FIBER
37 Barnstead |Thermolyne • 800-553-0039First Aid/Emergency ProceduresInhalation: REMOVE TO FRESH AIR. DRINK WATER TO CLEAR THROAT AND BLOW NOSE TO EVA
38Barnstead|ThermolyneOne Year Limited WarrantyBarnstead|Thermolyne Corporation warrants that if a product manufactured byBarnstead|Thermolyne and sol
39 Barnstead |Thermolyne • 800-553-0039
4IntroductionGeneral SpecificationsModels F30420CF30420C-70 & (-60)F30430CM & (-60)F30428CF30428C-70 & (-60)F30438CM & (-60)Overall Di
5 Barnstead |Thermolyne • 800-553-0039Typical Profile For Programmable Models Above. Other Profiles May Be Formed.4. The Type 30400 Automatic furnace
6Multi-Programmable Control (8 Ramp and 8 Dwell X 4 Programs)2 Ramp and 2 Dwell Programmable Control
7 Barnstead |Thermolyne • 800-553-0039Single Setpoint Controller
8
9 Barnstead |Thermolyne • 800-553-0039Unpacking1. Visually check for any physical damage to the shipping container. Inspect the equipment surfacesthat