Barnstead Nanopure™TOC -UFultrapure water systemOperation ManualSeries 1373 and 1381Model No. VoltageD11961 100V - 240VD12431 100V - 240VLT1373X1 • 11
Ultrafilter InstallationRemove the ultrafilter from its protective bag. Install theultrafilter as follows:1. Open the front (right side) door and righ
Other AccessoriesOptional accessories for this unit include a float switch,low pressure switch, N.I.S.T. Traceable CalibrationModule, standard remote
Installing the Control Panel in aRemote LocationFor your convenience, the control panel can be removedfrom the unit and mounted at a convenient locati
Water ConnectionsFeed Water Connection1. Locate the length of 3/8” O.D. tubing providedwith the unit with a quick disconnect insert onone end and a 3/
Tubing InstallationTo make tubing connections:1. Make sure the tubing is cut off reasonablysquare and that no plastic burrs or ridgesare present.2. *M
Main Power SwitchThe main power switch on the Nanopure TOC - UF islocated on the lower back right side of the unit (as youface the front of the unit),
16SwitchesWhen the main power switch (on the lower back right sideof the unit as you face the unit) is ON, the switches on thecontrol panel function a
17Cartridge packs will come bagged with four manifoldconnection caps.Part ApplicationDIamond Kit Organic Free R/O & Distilled FeedD50280 Ultra-Low
18Initial Rinse1. Attach the feed water line to the unit by snap-ping the quick disconnect coupling into the quickdisconnect body in the lower left ba
1911. Upon completion of the air purge, press the UPor DOWN arrow until “TOC Short Flush?” dis-plays. Press ENTER. (This flush will lastapproximately
Safety Information ...
208. Remove syringe luer cap and attach syringe tothe system luer fitting.9. Slowly inject solution into system and removesyringe. Avoid injecting air
211. Turn the system power ON by depressing the mainpower switch to the “I” position.2 The system greeting display indicates the type ofunit you own.
22Dispensing WaterUse the dispense knob for manual dispensing or, use the“Dispense” switch for volumetric or timed dispensing. See“User Settings.”Auto
23Adjusting Display Brightness1. From the (Idle) mode, when the display reads,“Nanopure (Idle) xx:xx:xx xx/xx/xx,” press the UP orDOWN arrow until dis
24Setting the Date and Time1. From the (Idle) mode, when the display reads,“Nanopure (Idle) xx:xx:xx xx/xx/xx,” press the UP orDOWN arrow until displa
25Setting the Cell ConstantThe actual temperature and cell constants are attached tothe cell cable internal to the Nanopure unit. This datahelps ensur
2613. After all digits have been entered, line 2 of thedisplay will read, “(NO) Finished.”14. If the cell constants displayed on line 1 of thedisplay
27USER SETTINGS3. Proceed through the options by pressingENTER until “Adjust Set Point” is displayed.4. Press the UP or DOWN arrow to choose “Yes.”5.
28Resetting the TOC Lamp TimerWhen the Nanopure TOC - UF is in an (Idle) position:1. From the (Idle) mode, when display reads,“Nanopure (Idle) xx:xx:x
Setting Volumetric Dispensing1. From the normal recirculating mode, when the dis-play is showing purity, press the UP or DOWN arrowuntil the display r
Your Thermo Scientific Barnstead Nanopure TOC - UFultrapure water system has been designed with function,reliability, and safety in mind. It is your r
30NotePurity displayed will have an asterisk(*) the the left of the reading (* xx.xMΩ-cm or * x.xx uS/cm) to inform theuser that uncompensated purity
Auto Standby(Factory set to “NO”)This option allows for a convenient method for the user toensure that the Nanopure TOC - UF unit will be placed inSta
324. Press the UP or DOWN arrow to choose “YES.”5. Press ENTER.6. Press BACK until you return to the (Idle) display.USER SETTINGS
33Feedwater Float or PressureSwitchAccessories D8964 (float switch) and D2706 (pressureswitch) are designed to protect the Nanopure TOC - UFpump by al
ACCESSORIESFigure 15: AY1367X1 Dispense Overflow Cutoff FloatNoteIf an auto dispense is stopped bythe AY1367X1, the Nanopure willchange modes to norma
Performing an ElectronicCalibration Using the OptionalN.I.S.T. Calibration ModuleIf you purchased the optional N.I.S.T. calibration module(Catalog No.
3612. Turn the main power switch to the OFF “O”position. Unplug the unit.13. Remove the calibration module and reconnectthe cell.14. Reattach the cove
37Connecting Nanopure toComputer and CommunicatingThrough the RS-232 Port UsingHyperterminal or ProcommHyperterminalRS-232 Capture Instructions1. Conn
38called capture.txt and the file will be an ASCII textfile.11. You should see the data on the computer screen.12. When you are finished storing data
12. Type in 21 <Enter>, sets to COM2: or (Type in20 <Enter>, sets to COM1:), depending on yourcomputer.13. Type in 24 <Enter>, saves
3. Avoid splashing cleaning solutions on clothing orskin.4. Ensure all piping connections are tight to avoidchemical leakage.5. Ensure adequate ventil
9. Press BACK twice to return to the (Idle) mode.10. During normal Recirculation mode, a purity andtemperature reading will be printed once every12 mi
41General Cleaning InstructionsDisconnect electrical service to the unit. Wipe exteriorsurfaces with lightly dampened cloth containing mild soapsoluti
42System CleaningFor UF units, it is best to perform cleaning proceduresearly in the afternoon. After cleaning is complete (approxi-mately 3.5 hours),
43and press ENTER. Display will now read, “UnitCleaning.”14. Unit may now be left unattended until thedisplay reads, “Cleaning Complete: PressENTER.”1
44Cartridge Pack ReplacementThe frequency with which you will need to clean your unitand replace your cartridge pack is dependent on yourfeed water’s
0.2 Micron Filter ReplacementReplace the 0.2 micron final filter whenever any of the fol-lowing conditions occur: the product water flow rate isreduce
467. Pull the ultrafilter out of the unit and turn it to theleft side. (This will make it easier to disconnectthe bottom fitting.)8. Using the tool su
47TOC Lamp Replacement1. Place unit in the “Idle” mode.2. Turn power off. Disconnect the power cord fromthe power entry module.3. Open right side door
485. Reattach the power cord.6. Operate normally.ShutdownIf the Nanopure TOC - UF is to be shut down for anextended period of time, the unit should be
49Problem Possible Causes SolutionsNanopure completely No electrical power to Nanopure. Ensure Nanopure power cord isinactive (pump and display connec
Congratulations on your purchase of a BarnsteadNanopure TOC - UF ultrapure water system. This waterpurification system is designed to provide pyrogen-
50Short cartridge pack life. Cartridge pack is beyond Check the expiration date. Cartridgeexpiration date. packs begin to lose capacity afterbeing sto
51Water is leaking between the Wing head screws are not Tighten wing head screws on unitmanifold and the cartridge tight enough. manifold.pack.New car
52Recommended SparesConsumablesConsumable parts are those required to support the day-to-day operation of this equipment. We establish twotypes of con
53Safety StockFor critical applications where performance with minimum downtime is required, we recommend that youmaintain a local stock of those part
54Unit/Meter/TOC Initialization& Meter Calibration CheckNANOpure DIamond (Idle)xx :xx :xx xx /xx /xxstart_stopkey pressedup or downkeys pressedbac
55Wiring DiagramDIAGRAM COMPONENT LISTREF.NO.DESCRIPTIONMODEL NO. AND OUR PART NO.(s)D11941 D11951BALLASTPC4C1 POWER ENTRY MODULE CEX183 CEX183CP1 PUM
56Ordering ProceduresPlease refer to the Specification Plate for the completemodel number, serial number, and series number whenrequesting service, re
Dimensions and Clearance RequirementsDimensions13.5” W x 19.5” H x 17.0” D (34.3 x 49.5 x 43.2 cm)ClearancesSides - 9” (22.9 cm) minimum for servicing
Two Year Limited WarrantyThis Thermo Scientific product is warranted to be free of defects in materials and workmanship for two (2)years from the firs
Electrical RequirementsThe Nanopure TOC - UF is equipped with 2 power cords to be plugged into an electrical outlet of the appro-priate voltage.U.K. c
1. Remove the unit from its shipping container.Remove all contents carefully. Ensure that theultrafilter, feed and drain tubing, cleaningcartridge, wa
9Choosing a SiteThe Nanopure TOC - UF features a removable controlpanel display which allows the system to be mountedalmost anywhere within the labora