Title
Barnstead
Thermolyne
NANOpure® DIamond™
Ultrapure Water System
Installation/Operation
Qualification
Protocol
BT.001.01.A
Customer
Protocol
____________
Revision
0
Customer
Revision
____________
Equipment ID
NANOpure®
DIamond™
Ultrapure
Water
System 1
Customer
____________
____________
____________
____________
____________
BT.001.01.A
09262000 Page 15 of 21
B. Operational qualification
Step Test or Control
Device
Specification or Acceptance
Criteria
Actual Result
1 Power switch NANOpure
Diamond™ Ultrapure
Water System is receiving power.
Control panel LCD is lit.
Y N
Y N
2 Configuration Greeting displays correct model
when power switch is activated.
Analytical
UV
UF
UV/UF
Y N N/A
Y N N/A
Y N N/A
Y N N/A
3Normal
Operation
Does pressing the Start/Stop key
start the pump? Unit displays
***MΩ-cm
Does pressing Start/Stop key again
stop pump? (Unit returns to IDLE)
Press the Start/Stop key to start unit.
Does pressing Standby key stop
pump, and display “Standby?”
Y N
Y N
Y N