Title
Barnstead
Thermolyne
NANOpure® DIamond™
Ultrapure Water System
Installation/Operation
Qualification
Protocol
BT.001.01.A
Customer
Protocol
____________
Revision
0
Customer
Revision
____________
Equipment ID
NANOpure®
DIamond™
Ultrapure
Water
System 1
Customer
____________
____________
____________
____________
____________
BT.001.01.A
09262000 Page 9 of 21
Unit Location/
Installation
Continued
Has initial rinse-up been performed?
If not, see manual for initial rinse
instructions.
Has the final 0.2 micron filter been
installed?
Y N
Y N
2 Feed Water Supplied feed water must be
reverse osmosis, deionized or
distilled,
<70 ppm total dissolved solids (TDS).
Record actual TDS.
What is the TOC (total organic
carbon) level of the feed water
(ppm) provided to the unit?
(required value is <1.0 ppm)
Record actual TOC
What is the turbidity level of the
feed water (as measured in NTU)
provided to the unit?
(required value is
<
=1.0 NTU)
Record actual turbidity
What is the pressure of the feed
water (as measured in psig)
provided to the unit?
(required value is <=100 psig)
Record actual pressure
__________ ppm
___________ ppm
___________ NTU
___________ psig